Drop-down menu navigation.
|

|
Constant background monitoring.
|

|
Sleep Mode for extended recordings.
|

|
Greater Math Functions in MAS™.
|

|
User adjustable span/offset settings.
|

|
Event/Time-based triggerd recordings.
|

|
Folder-based file management.
|

|
Long-life Nicad battery pack.
|

|
24 bit A/D converter.
|

|
ProFlash upgradeable firmware.
|

|
MAS Windows based software.
|

|
CE approved.
|

|
512 Kilobyte non volatile memory. (Approx 400,000 samples)
|

|
4.3"x2.3" (100mmx60mm) LCD screen.
|

|
Long-life Nicad battery pack.
|

|
RS232 Interface with cable.
|

|
3 Year Warranty (on instrument only).
|
ITEM
|
SPEC
|
REMARKS
|
|
Model: MMS3000-GP4 & GP8 4 & 8 Current Loop Input Channels with General Purpose Converstion Software
|
|
CURRENT RANGE
|
Nominal current range
|
4 to 20mA
|
4mA corresponds to 0 units (units selectable in software), 20mA corresponds to full-scale indication.
|
Maximum current range
|
3.2 to 24mA
|
Offset and span are user-adjustable.
|
CURRENT RESOLUTION
|
|
2µA
(1 part in 2,000 @ 4mA, 1 part in 10,000 @ 20mA)
|
|
CURRENT ACCURACY
|
|
?0.1% of reading ?5µA
|
Offset and span calibrated to nominal values. Instrument temperature 73 ?18°F
(23 ?10°C).
|
ENGINEERING UNITS
|
Units
|
Metric and Imperial units for temperature, displacement, force, pressure, voltage, current, plus percentage.
|
User can create custom Engineering Units.
|
Scaling
|
pico, nano, micro, milli, kilo, mega, giga, tera
|
Units and scaling are user-configurable.
|
MINIMUM SAMPLING INTERVAL
|
|
Number of channels selected (User-selectable)
|
Minimum sampling time (Seconds)
|
GP4
|
GP8
|
|
1
|
1
|
0.2
|
2
|
2
|
0.3
|
3
|
3
|
0.5
|
4
|
4
|
0.6
|
|
5
|
0.7
|
|
6
|
0.8
|
|
7
|
0.9
|
|
8
|
1.0
|
|
Minimum time to scan all selected channels, not each channel.
Sampling interval is programmable from 0.2 seconds to 60 hours.
Model GP4 has 4 channels only.
|
DATA LOGGING
|
Data Storage Format
|
Up to 50 named folders
|
User-specified name entered from keypad.
|
|
Multiple files per folders
|
Each file has unique time and date stamp.
|
|
Up to 59,999 data points per file
|
MAS plots limited to 15,000 samples per channel.
|
Data Storage
|
512 kilobytes ?approximately 400,000 individual samples
|
Non-volatile memory (with battery backup).
|
Folder Name
|
Up to 16 alphanumeric characters
|
Number of samples depends on number of folders and files.
|
DISPLAY
|
|
LCD with graphic capabilities
|
|
Resolution
|
240 x 128 pixels
|
|
Viewing Area
|
4.3" x 2.3" (110mm x 60mm)
|
|
Backlight
|
Electro-luminescent
|
|
PROFLASH
|
|
Allows internal firmware to be upgraded via built-in serial port
|
Download firmware upgrades via the Internet
|
COMMUNICATIONS
|
|
RS232 - 15kV ESD Protected.
|
Cable with DB9 connector included
|
Baud Rate
|
9600, 19200, 57600 bits per second
|
Automatic Baud Rate Selection
|
INPUT TERMINATION
|
|
Screw Terminals, 5mm pitch, 2.5mm wire size
|
4 pairs for model GP4, 8 pairs for model GP8.
|
LOOP BURDEN
|
|
Maximum 6 Volts at 20mA loop current
|
|
ISOLATION
|
|
Optically Isolated
|
Between each current loop input and between each current input and instrument outputs e.g. RS232 port. The instrument is not intended for connection to "live" surfaces (voltage exceeding 50 VDC or 32 VAC).
|
BATTERY
|
Type
|
Custom Nickel-Cadmium pack
|
|
Voltage
|
7.2 Volts nominal
|
|
Capacity
|
1500 mAh nominal
|
|
Operating Time
|
13 hrs with backlight off, 7 hrs backlight on - nominal
|
|
CHARGER AND CONDITIONER
|
|
Built-in dual rate charger ?Automatic and Manual control
|
AC Adapter included (13.5 ?1.5 VDC, 1A output)
|
Fast Charge Rate
|
0.7A nominal
|
2.5 hours for complete charge nominal
|
Discharge Rate
|
0.5A nominal
|
Combats NiCad battery memory effect
|
MECHANICAL
|
Size
|
9.7"W x 6.1"L x 3.0"H (247mm x 154mm x 75mm)
|
Including protective boot
|
Weight
|
4.4lb (2kg)
|
Including protective boot and strap
|
ENVIRONMENTAL
|
Temperature/Humidity
|
|
|
Operating
|
32°F
to 122°F
(0°C
to 50°C) / 70% RH
|
Non-condensing
|
|
32°F
to 86°F
(0°C
to 30°C) / 80% RH
|
Non-condensing
|
Storage
|
14°F
to 140°F
(-10°C
to 60°C
) / 95% RH
|
|
EMC
|
EN50081-1
|
Radiated and conducted emissions
|
|
EN50082-1
|
ESD Immunity and RF field immunity (EMI).
|