Model
CAL10
Thermocouple Calibrator
Measures Thermocouples, Voltage and Current, Programmable





Dimensional Enlargement

Calibrate electronic process instruments such as controllers, transmitters, transducers or recorders with model CAL10 Thermocouple Calibrator. Model CAL10 is designed to independently measure and simulate dc mV or V and common thermocouple signals (Type J, K, T, E, J (DIN), and T (DIN) or measure current. Input/output can be operated simulanteously with independent range selection

A multitude of programmable functions, such as square root extraction for flow signals, ramp/step increment value, selectable engineering units, isolated signal conversion or automatic changing of the output value simplify many calibration tasks. Calibration values can be stored and recalled using the Low, Mid, or High keys. Cold junction compensation can be selected for automatic or manual. Program the automatic shut-off to conserve battery life.

The Model CAL10 calibrator features 500 VDC input/output isolation and reverse polarity protection. View input and output valvues on the 24-character LCD. The unit includes three 9V batteries, two sets of dual test leads, a dual binding post adapter, carrying case, and instruction manual.

PHYSICAL DATA
    Input Ranges: Voltage: -10 to 24 mV DC/ -10 to 120 mV DC (autoranging), 0 to 12 VDC; Current: 0 to 24 mA DC; Temperature: J: -346 to 2192°F (-210 to 1200°C); K: -420 to 2500°F (-250 to 1370°C); T: -400 to 752°F (-240 to 395°C); J (DIN): -328 to 1652°F (-200 to 900 °C); E: -420 to 1832 °F (-250 to 1000°C); T (DIN): -328 to 1112°F (-200 to 600°C).

    Accuracy: Voltage and Current: ±0.025% of range; Temperature: see table below.

    Repeatability: Better than 0.03% of range.

    Resolution: 0.1°C/F

    Warm-up Time: 60 seconds, maximum.

    Input Impedance: 1 megohm minimum plus 10 nanoAmps maximum for V, mV and T/C inputs; 40 Ohms for mA DC.

    Output Ranges: 0 to 120.0 mV DC (2K Ohms min. load); 0 to 12.0 VDC (2K Ohms min. load); Thermocouple Types J, K, T, E, J (DIN), and T (DIN) similiar to input ranges.

    Response Time: 0.5 seconds.

    Input/Output Isolation: 500 VDC.

    Repeatability: Better than 0.03% of range.

    Operating Temperature: -4 to 113°F (-20 to 45°C).

    Display: 24-character LCD, 3/16" (5 mm) character height.

    Indication: Up to 5 digits depending on range.

    Electrical Connections: Standard banana jacks on 3/4" (19 mm) centers.

    Power: Three 9V alkaline (included).

    Battery Life: 100 hours, approximate.

    Weight: 1.15 lbs (0.52 kg).

Thermocouple Range Accuracy
J     Fe(+) -250 to 2192°F ±1°F
    Cu-Ni(-) -346 to -250°F ±2°F
K     Ni-Cr(+) -350 to 2500°F ±1°F
    Ni-Al -380 to -350°F ±3°F
-420 to -380°F ±5°F
T     Cu(+) -100 to 1112°F ±1°F
    Cu-Ni(-) -300 to -100°F ±3°F
    Cu-Ni(-) -400 to -300°F ±5°F
E     Ni-Cr(+) -400 to 1832°F ±1°F
    Cu-Ni(-) -420 to -400°F ±2°F


STOCKED MODELS
Model CAL10 Thermocouple Calibrator